Screening and Scanning EM Film


Procedure with screen shots available on web site http://www.nysbc.org/facilities/CEM/Zeiss_PhotoScan_Usage.html

Download MSWord format with screen shots at end * Zeiss-Photoscan-Usage.doc:

Contents

General Guidelines

  • Always use gloves
  • Always have something between the glass plattens to prevent scratching
  • Never use compressed air sources (which inevitably contain water or oil),
    • use brushes to remove dust or clean glass while wet with glass cleaner - experts only
  • Transfer and Delete your images from the local disk immediately after scanning they won't be there when you return
  • Turn off scanner when no foreseable use later that day (to preserve lamp hours)

Turning on Scanner

  • Scanner should be left on, with the light bulb off
    • The on switch of the bulb is on the right side of the scanner, toward the back. Up/down = on/off
    • Turn the scanner off then on again using the main on/off switch on the left side of the scanner
    • Turn the bulb to the on position
    • Computer will beep a few times, then the scanner will do a self-test
    • Wait until self test has finished "Self Test OK"
    • If no beep and no self test initiated, please call for help
  • Turn on Windows Computer (open black door underneath scanner and toggle power switch)
    • Note that "Carl Zeiss" SCSI device is identified during start up
  • Log onto computer
  • Open Hyperterminal window (Controller Hyperterminal icon on Desktop - should open automatically)
  • Start the Zeiss Photoscan software
    • Should get a bunch of messages from the scanner in the Hyperterminal window

Loading the film

  • Use gloves
  • Open cover
  • Raise upper glass platten, which can be held up by the cover
  • Remove dust from film and glass with brush
    • Do not use glass cleaner without assistance from NYSBC staf)
  • Place film on lower glass platten emulsion down
    • Cutout corner goes on the upper left
  • Lower upper glass platten
    • Glass will break if cover is lowered first
  • Lower cover

Starting program

  • Open Photoscan program, if not already started
  • Select File -> New Scan
  • Scan Setup - select nysbc_def.par
  • Save setup parameters in your own file by changing nysbc_def.par and hitting save

Preview film

  • Select file name on "Scan Setup" tab (e.g., E:\stokes\test-mrcgrid.tif)
  • Select course Resolution on "Scan Setup" tab (e.g., 28 um)
  • Make sure "Overview" is set for "None"
  • Should not need to change anything on "Camera Definition" tab
  • Select wide range for OD on "Film/Output" tab (e.g., 2.0 to 0.2)
  • Save setup parameters (program crashes periodically)
  • Select Scan -> Scan to Disk from main menu (F7)

Check normalization data

  • Upon initiation of first scan of the day, a normalization is automatically done.
  • Look at Hyperterminal window specifically for the following data * Information appears on monitor on top of scanner. Watch closely as it will scroll off screen.

No. of ff-values: B 0 G 0 R 0
Minimum-values: B 205 G 215 R 209
Maximum-values: B 245 G 251 R 250
Measured with Exposure Time: 2.10 ms ok.

  • Min values should be within 50-60 of max values
  • Max values should be < 255
  • Three channels should be within 20 counts of one another

No. of 0-values: B 0 G 0 R 0
Minimum-values: B 3 G 2 R 3
Maximum-values: B 7 G 4 R 5

  • Max values should be < 10 with no zeros

No. of gain-values cut to 2047: B 0 G 0 R 0
No. of gain-values cut to 0: B 0 G 0 R 0

No. of offs-values cut to 2047: B 0 G 0 R 0
No. of offs-values cut to -2048: B 0 G 0 R 0

  • Number of values cut should all be zero
  • Any anomalies may indicate time to change the light bulb

Select Scan Area

  • Use "Define Scan Area" tool along top tool bar (or in "View" menu)
  • Adjust size of box to that desired for final scan
  • Select File -> Scan Properties
    • check Histogram tab and adjust OD range under Film/Output
    • adjust Resolution (e.g. 7 um)
    • adjust image size (select "pixels" and enter size)
  • Save setup parameters
    • please use your own file
    • program will be crashing soon, but parameters will be saved

Final Scan

  • Select Scan -> Scan to Disk (or hit F7)
  • Select File -> Scan Properties and look at Histogram tab
  • Adjust any parameters if you want to rescan the image
    • OD range usually adjusted so that the densities comfortably fit within the chosen OD window.
    • However, at 7 um scan, min OD >0.6 is likely to cause an error (scanning hangs at the end of the 1st swath). If so, wait 3-4 minutes for a SCSI timeout. Lower the min OD value to 0.5 and rescan. This does not appear to be a problem with 14 um scan.
  • Save the setup parameter file (please use your own file)
  • Close window with image and agree to save the corresponding file
    • (otherwise will be deleted)

Shutdown

  • Exit Photoscan
  • Remove film from scanner
  • Replace rubberized insert between glass plates (brush off any dust first)
  • Be mindful of upper glass platten before closing lid
  • Shutdown computer
    • You can leave computer on if returning in foreseeable future.
    • However, SCSI connection between computer and scanner seems to get corrupted after significant periods of time.
  • Turn off scanner light bulb (right side of scanner, towards back
    • Leave scanner itself on, as it sometimes has problems starting from a cold-state

Burn CD

  • Open door to lower computer (black door under scanner)
  • Insert blank CD in the drive
  • Open "Nero Express" on desktop
  • Select Data -> Data Disk and select files to include on CD

Transfer files to File Server (dl380)

  • Run the sftp software (icon on Desktop)

Transfer files to USB stick

  • Connection is via an extended USB cable
  • Speed is slow (USB1) so be patient

  • Open "My Documents" on desktop or "Windows Explorer" from Start Menu
  • Hit "Search", select "Computers", enter "dl380"
  • Locate your home directory (select user name twice for your home directory)
  • Open second "My Documents" and locate your scanned files on the local disk
    • (e.g., E:\stokes\*.tif)
  • Copy and Paste from CD to dl380 file server

Troubleshooting

  • Call staff for any problems starting or completing scans
  • If you cannot log in to dl380, just burn a CD
    • Contact staff if blank CD's are not available in scanner room
  • Photoscan software crash:
    • In event of a crash, try restarting Photoscan program
    • If simple restart doesn't work, try logging out of Windows PC and logging back in
    • If logout/login doesn't work, try rebooting Windows PC
    • If you still cannot connect, turn off Windows PC, turn off scanner, and restart from scratch
  • Zeiss scanner errors
    • Please contact staff if scans fail to complete or normalization fails

Testing Scanner