Protocols - Using the PDS 1010GM Microdensitometer
Download MSWord file with these instructions: PDS-1010GM-Usage.doc
To start
- Put on gloves
- Sweep glass with anti-static brush
- do not use canned air on glass
- Clean film with quick burst of canned air
- Place film on glass emulsion side up
- Plug in vacuum pump for film holder
- pump located on other side of the wall but plug comes through the wall
PDS Alignment
- Select "7" on inner filter wheel (protects PMT)
- Turn on Power, Lamp, PMT and allow to warm up
- Select upper and lower apertures and lenses depending on required pixel size
- i.e. For 13 um scan:
- Use "B" aperture" and "2" lens both above and below the film
- For 6.5 um scan
- Use "B" aperture and "4" lens above and below the film
- Move to clear area on film
- Focus lower lens until see sharp edges of lower aperture on screen
- Remove aperture with lower left dial
- Focus upper lens until film grain is sharp (not lamp fiber optics!)
- Replace aperture
- Center upper lens roughly on main screen- with 4 set screws
- Check center with monocular (upper aperture should be inside lower aperture)
- Set gain:
- Select "C" on filter wheels (both)
- Toggle to TRANS, CALIB to 1
- adjust PMT voltage -> 1.00
- Toggle to DENS, CALIB to N
- Leave in DENS - N mode
- Select "7" on inner filter wheel for safety - especially with small apertures
- Move to center of desired scan
- Reset X-Y positions with red push buttons
- Remove aperture with lower dial
- Check lens focus on film grain using monocular eyepiece
- Replace aperture
- Select "C" on filter wheel
- Push SELECT -> AUTO
Computer Controls
- Put computer in MSDOS mode
- Either choose MSDOS mode upon startup
- Or choose "Shutdown" and specify "Start in MSDOS mode"
- Start SCANS-A-LOT control program
- look over configuration, press ENTER
- Enter "ED" to edit scan
- "2" to change name of file
- "3" to change scan parameters and speed
- "ctrl + Z" to go back to edit menu
- "0" to exit out of edit menu
- "h" to display Scans-A-Lot prompts
- "CS" to begin scan
- "Q" to exit to windows
Finish scan
- Turn off PMT, Lamp, Power and recover PDS when finished
Transfer files to dl380
- start Windows
- type "win" at MSDOS command prompt
- click winscp361.exe
- Copy files from c:\pds on PDS computer to dl380 /home/your file folder
- Delete files from PDS computer after confirmed transfer