Interested in EM?
Using the NYSBC
Principles & Protocols
NYSBC Equipment
Seminars & Courses
Publications associated with CEM at NYSBC
Jobs Available


Microscope Schedule
Logbook
NYSBC Intranet
NYSBC Contacts
Cryo-EM home



NYSBC home
NMR Facility
X-Ray Facility
Protein Production Facility
Electron Crystallography Center


Cryo-EM Site map

JEOL 2100 Instructions


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM


Initial Start-up

  • "Image shift Power up" should be on task bar ALWAYS!!!! * "High Power" Selected * Found in Start Menu
  • "JEOLTEM" should be running ALWAYS
  • Do NOT shutdown any computers
  • QuickCapture shortcut will show column vacuum gauge
  • Do not open beam valve until column vacuum < 2X10-5 Pa
  • Make sure "OL" button on left panel is lit
    • Found under "Aperture Control"

Start-up

  • Fill LN2 dewar at least 20min prior to use
  • Top up again after 20 min - then will last 1/2 day

Check beam before inserting sample

  • Insert Dummy holder
  • Open "Beam" valve
  • If no beam lower magnification
  • Insert and center condenser aperture
    • Suggest spot size 5 and smallest C2 aperture that gives you a bright enough beam
    • At 50kx, go to crossover, center spot with beam shift, expand beam by over focusing C2, center illuminated area

Loading Room Temperature Grid

  • Use special pin to lift spring-loaded latch
  • Lift and remove specimen tip from holder
  • Place specimen tip in nylon block - engage pin into matching hole
  • Unscrew latch slightly (~1 turn) and pivot to one side
  • Drop grid into position
  • Pivot latch back into position and tighten screws (DO NOT OVER-TIGHTEN)
  • Reinsert specimen tip into holder - make sure to engage pin - tip should be flush against holder and held firmly

Sample Insertion

  • Make sure the stage is neutralized before inserting sample
  • Open Vacuum System panel (under Monitor)
  • Put toggle switch to "pump"
  • Insert specimen rod to first stop - provide slight counter-clockwise torque to ensure interaction with relay inside
  • Wait until Specimen/PiG4 pressure in "Vacuum System" panel reaches 39 for cryo - 38 or lower for rm temp
  • As of July 2007, PiG4? comes down much faster. On first loading of RT sample, wait until it reaches 29, otherwise vacuum takes a long time to recover
  • Rotate holder clockwise 5 deg, insert, continue to rotate clockwise until holder enters vacuum
  • Can leave toggle on either pump or air - position should be monitored whenever removing/inserting holder
  • Under Properties-->Stage, choose the type of holder you inserted (standard tilt, high tilt, or cryo)

sampleinsertion.jpg

Aperture Insertion

  • Insert and center condenser aperture
    • Suggest spot size 5 and smallest C2 aperture that gives you enough beam strength
    • At 50kx, go to crossover, center spot with beam shift, expand beam by over-focusing C2, center illuminated area
  • Insert and center objective aperture
    • Use "OL" aperture, 1st cutoff at ~2.1A, 2nd cutoff at ~3.3A
    • Go to SA diff and focus diffraction pattern using DIFF FOCUS
    • Turn left panel 90 deg clockwise to aid centering (right/left correct, up/down reversed)

Basic Alignment procedure

  • Open JEOLS hex display window
  • Open Maintenance alignment panel
  • Press Std Focus to make sure you are not doing alignment at some crazy focus

Initial alignment

  • Insert sample or dummy. Sample needed for Voltage Center alignment.
  • Insert and center condenser aperture 1 or 2
  • Open column valve, go to 40-50 K mag
  • Select desired alpha setting (2 or 3) (2 for iMDS; 3 for SerialEM)
  • Press Bright tilt button
  • Condense beam and move to center with Bright Shift
  • Adjust condenser stigmation to make beam symmetric
  • Go to Eucentric position
    • Technique 1
      1. Move stage to area where you can see something
      2. Press Eucentric Focus button
      3. Open Alignment panel
      4. Turn on both image wobblers (x and y)
      5. Adjust specimen Z position until image movement minimizes
    • Technique 2
      1. Go to ~ 5K mag
      2. Condense beam
      3. Should see a bright spot surrounded by halo
      4. Adjust Z height to minimize halo
    • Technique 3
      1. Go to moderate mag (10-20K)
      2. Center feature of interest using stage shift
      3. Tilt stage to some angle (5 degrees or less to start)
      4. Adjust Z height until feature of interest is re-centered
      5. Repeat for higher tilts, up to 15-20 degrees. Do positive and negative tilts
  • Go to high mag (~300K) and condense beam
  • Press Bright Tilt
  • Put Caustic spot in center using lower def/stig xy knobs

Align gun shift:

    1. Go to very high mag (300K)
    2. Open Alignment Panel
    3. Go to spot 5, condense beam and center with Bright Shift
    4. Go to spot 1
    5. In Alignment Panel, click on Gun (or hit F3)
    6. Move beam back to center with Gun Shift
    7. Go to spot 5, click on Bright Shift
    8. Move beam back to center with Bright Shift
    9. Repeat above steps until alignment converges

Align Gun tilt

    1. Go to spot 1
    2. Choose Gun under Alignment Panel
    3. Condense beam
    4. Turn on anode wobbler
    5. Adjust Gun tilt until beam expands and contracts about same center
    6. Turn off anode wobbler
    7. Re-center beam with Gun Shift
    8. go to spot 5 -- beam should not move much at this point
    9. Press Bright Tilt button to be safely out of Gun shift mode

Align Pivot points

  • Go to desired spot size (5) and adjust condenser astigmatism
    • Tilt
      1. Condense beam
      2. Under Alignment panel, press tilt
      3. Press tlt-x, adjust tilt with def/stig x until movement is minimized
      4. De-select tlt-x, select tlt-y and adjust with def/stig y until movement is minimized
    • Shift ( STAFF ONLY )
      1. In mag mode, fully expand beam, then go to SA DIFF
      2. Fully expand beam using Brightness, then condense beam to caustic spot using Diff Focus
      3. Under Alignment panel, choose Shift X and select Shift compensator
      4. Using Def/Stig X, minimize movement of spot
      5. Repeat for Shift Y

Align Voltage Center

    1. Specimen inserted, eucentric, and focused
    2. Align feature of interest with stage shift
    3. Turn on HT wobbler in Alignment panel
    4. Press Bright tilt
    5. Adjust bright tilt with def/stig x y until center of expansion is at center of image
    6. You can use the off-axis camera if this is easier -- first shift the image to off-axis position with PLA

Repeat

    1. After above steps completed, repeat 2 or 3 times until alignment no longer changes

  • When done, save alignment in case something (MDS, iMDS, SerialEM) messes it all up again

iMDS setup

  • DO NOT USE MDS use iMDS

  • Search and Focus modes = PL Align
  • Photoset mode = Beam Shift
  • In general, when setting up, it is better to change one thing at a time. Cycle through the modes to stabilize, then move on to the next thing.

  • Set microscope to conditions desired for imaging:
    • Microscope aligned for the day at alpha 2
    • Condenser lens inserted and centered
    • Objective lens inserted and centered
    • Desired magnification for recording images
    • Desired beam intensity for recording images

  • Open iMDS panel
    • Select "Yes" in initial dialog box
    • Current microscope settings will now be copied into all modes
  • Set up Search mode
    • Mag mode 1 (NOT SADIFF)
      • Set mag to 10K, and fully expand beam
      • Go to SADIFF mode
      • Choose combination of camera length and Diff Focus to get correct mag and brightness (should just be short of covering CCD with 2nd objective aperture inserted)
      • Adjust position of beam with PLA (NOT Image shift!!) to put beam over Fast scan camera

  • Set up Focus mode
    • Got to Focus in iMDS
    • Set up mag/beam to desired level (200 K, spot 5, 10 pA/cm2 is usual).
    • Adjust position of beam away from Photo-set position by clicking on desired spot in diagram (2 um away should be good)
    • Yellow line on diagram is tilt axis
imds-map.png

      • Beam will not be centered but should still be on large screen
      • Cycle a few times then center beam over Fast scan camera using PLA align

  • Set up MDS Photoset mode:
    • Go to MDS Photoset in iMDS
    • May need to re-center beam with Bright Shift
    • Whenever beam is re-centered, Focus beam will likely also need to be moved with PLA
  • Alignment of Search and Photoset
    • Go to Search mode
    • Find a clean area of ice to burn, or an easily identifiable featire
    • Go to Focus, then Photoset. Burn a nice hole using a long exposure with the screen down
      • Go back to Search. Center burn hole using PLA shift (NOT image shift)
  • Optional lowering of beam intensity in search (recommended for cryo)
    • in Search mode, open the Custom tab
imds-custom.png

    • Note the hex values of spot size and alpha selector
      • Spot 5: 9860 Alpha 2: 9510
    • Click Custom on
    • Change the hex value of Spot size to a high number (FFFF is the max)
    • Go to Focus
    • Spot and alpha are likely reset to zero. Check by clicking up/down arrows on value
    • If zero, replace with the original numbers corresponding to spot size and alpha
    • Go to MDS Photoset and do the same
    • Cycle a few times then recenter MDS Photoset with beam shift, then recenter Search and Focus with PLA

LowDose in Serial EM

Holder Removal

  • IMPORTANT FIRST STEPS:
    • Change toggle on goniometer from pump to air
    • Close Beam Valve
    • Monitor Vacuum (Vacuum overview and Quickcam gauge)
    • Double Click "Stage Neutralize" button on top panel
  • Gently pull out holder to first stop
  • Turn counterclockwise until 2nd stop
  • Pull out to 3rd stop (short distance).
  • Listen for three clicks before removing holder. V36 valve will turn green - nitrogen is pumping into goniometer to vent holder.
  • On the Valve status page, you will see valves V21, V34 and V36 light up on sequence. N2 vent is done when valve V36 goes black again.
  • Holder should be easily removed - no vacuum breakage like on the Tecnai.
  • If a mistake is made and you break vacuum, or you forgot to change the toggle, DO NOT INSERT ANOTHER HOLDER. Call Ruben, KD, or Bill for help.

End of day procedure

  • After removing holder, the dummy holder should be inserted. Keeping holder in microscope helps to protect vacuum
  • return microscope to standard condition
    • lower main viewing screen
    • Select "All off" on the "Free Lens Control" panel
    • Turn off "MDS"
    • Set mag to 40kx
    • Spread beam to size of screen
    • Neutralize stage
    • replace cover on viewing chamber
    • close beam valve
    • close EM Menu/!SerialEM
    • turn off monitors
    • change film
    • open panel just below viewing chamber
    • turn handle 90 deg clockwise
    • wait ~1-2 min and door will pop open
    • take both film boxes into dark room
    • replace film boxes (supply box in desiccator receiving box behind microscope)
      • check that film box lids are engaged in matching slots
    • slide tray into place
    • swing door closed - give last push and turn handle 90 deg counterclockwise
    • close panel
    • develop film and replace fresh unexposed film in dessication, empty receiving box behind microscope
      • activate spring in supply box so film pushes against top and prevents accidental opening
    • Start ACD & Bake (under Maintenance)
    • put heater element into liquid nitrogen dewar (nitrogen will blow off) and make sure it is plugged in properly
    • remove C2 and objective apertures (align red dot to black dot on microscope) and sample
    • turn ACD Heat "ON"

Shut Down of instrument

  • This procedure is to be used only under dire conditions, such as a power failure.
  • Open the HT window from the microscope GUI and put the FEG in Stand By mode
  • Expand this window ("more" at the bottom right) and press the button for HT OFF
  • When this finishes (~4min), turn off the power of the microscope (EM Stop) at the panel in the bottom-left to the microscope.

How to bring the microscope back to life

  • This should be done by a staff member only!!!
  1. Make sure there is water flow to the microscope (this is off if the reason for the shutdown was to bake the instrument)
  2. Push the "lens" button on the microscope panel (bottom-left panel)
  3. Rotate vacuum gauges to the proper reading ranges (10^-6) at the main panel in the utility room. Make sure the vacuum reading for the gun is at an acceptable value (<2x10^-6)
  4. Turn the HT on in the microscope GUI (might need to set an arbitrary value for this, which is done to be able to carry out the next step)
  5. Open and close the gun valve a few times (~3)
  6. Turn the HT off (it takes some time to do so and gives a message not to touch the conditioning handle)
  7. Climb up the ladder and turn the gun knob to conditioning mode
  8. With the HT reading 0kV, turn the HT on (i.e., just press the friggin button)
  9. Turn manual conditioning on
  10. Click the HT increment to 100kV
  11. In auto HT, set the target to 160kV, change the steps to the smallest possible step, and adjust the time between steps so that it will take 30min to complete
  12. Set target to 180kV and adjust the time span so that it is done again in ~30minutes
  13. Do the same from 180kV to 200kV and let it sit at 200kV for 10 minutes. It is important to make sure that it only stays at 200kV for 10 minutes at this stage, since the actual voltage is 205kV at conditioning.
  14. Turn the HT off in the HT manual button (not on the one for HT and Emission). There will be a warning window. When it disappears, climb up the ladder again, and set the cup to "operate"
  15. Hit the blue button to auto start the HT and gun. This will take a LONG time
  16. Make sure that the cameras are on, and when they reach a stable temperature, do the flat fields.
  17. Go for a coffee.

JeoL2100Controls

JeoL2100Problems

  • Expert users only, please contact Bill.
  • Restarting the microscope controls

SerialEM Calibrations

  • done 08-07-07
  • be sure to use alpha 3 in SerialEM, otherwise beam movement adjustments won't work

-- KdDerr - 06 Sep 2006