F20 Daily Checklist
- Cool microscope cold finger
- Insert holder to first stop
- Open colum valves
- Start Low Dose and click through systematically to "exposure" mode
- Load Lose Dose file you will be using
Operation Parameters for alignments
- 3850-3950 extraction voltage
- Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.)
- 50 micron condenser aperture (check white board, for now it's position #2)
Gun tilt (check)
- Magnification 10kx
- Check exposure time without the small screen
- Around 2 sec is appropriate
- 2 sec exposure means 1 nanoAmp of current
- if the exposure time is not close to 2 seconds double check for correct operation parameters
- If the operation parameters are correct adjust Gun Tilt
- Location: tune/alignments/gun
- Adjust the gun tilt alignment to locate the brightest spot
- After adjusting Gun tilt:
- If exposure time is over 2 sec increase the extraction voltage and readjust gun tilt
- If exposure time is less than 1.4 sec lower the extraction voltage and readjust gun tilt
Condensor Stigmation
- Spot size 2
- Magnification 250kx
- Condense beam to 30mm size
- Press stigmator button
- Use multifunction knobs to make beam round
Check image of FEG Tip
- Condense beam to point
- Beam may not be round at crossover, this is the image of the FEG tip
- 5mm circle = 10nm
- You want the majority of your beam to fit within the 5mm circle
Flatfield Tietz camera